Entering Settings
All autosampler settings can be entered in the Autosampler Settings screen.
You can access this screen by selecting the Settings button [F1] in the Analysis
screen. From here you can access the timetable function (to set up a timetable
for the Autosampler only), the runtimes screen and the injector program
screen. Commonly-used settings such as injection volume and vial number can
be set directly in the Analysis screen. More specific settings (draw speed, eject
speed, draw position offset) can be entered using the More... button [F1] and
choosing the Settings option.
Setting the Injection Mode
There are two modes which can be set:
• Standard
• Injector Program
1200 Series Handheld Control Module Users's Guide
•
without pre-defined needle wash,
•
with pre-defined needle wash (wash vial number can be set).
If the pre-defined needle wash function is enabled, the needle moves prior
to the injection (with reagent in the needle) into the wash vial to remove
sample deposits from the needle outside. This prevents sample stay in the
seat and create memory effects on the next injection cycle.
See
"Running an Injector Program"
Using the Autosampler
Entering Settings
on page 149.
5
77