Running Multiple-Vial Analyses
Multiple-Level Calibration Sequences
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence button [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
3 Select the Calibration button [F1] to display the Calibration Settings screen for
4 Enter the following information:
5 Select the Done button [F6] to accept entries.
6 Move the highlighted bar to line 2 (by pressing the Selection key down) and
7 For sequence line 2 enter:
8 Select the Calibration button [F1] to display the Calibration Settings screen for
142
Vial Range
7 to 12
#Inj.
1
Inj. Volume
Default
Method
METH1
sequence line 1 and choose Edit.
Calibration vial range 1 to 3
Number of injections 2
Inj. Volume
Default
Calibration method
METH1
Recalibrate every
2 vials multi
Before
On
After
Off
press Enter.
Vial Range
13 to 21
#Inj.
1
Inj. Volume
Default
Method
METH2
sequence line 2 and choose Edit.
1200 Series Handheld Control Module Users's Guide