Recalibrating With Multiple Groups of Standards

There are two different types of samples, A and B that need to be analyzed.
The analysis for sample type A requires a 5 µl injection and a stoptime of 8
minutes.
The analysis of sample type B requires a 2 µl injection and a stoptime of 5
minutes.
For sample type A:
• there are 3 calibration standards of different concentrations and 6 samples,
• each sample must be analyzed once,
• the calibration standards must be analyzed in duplicate and re-analyzed
• The calibration standards of type A are in positions 1, 2 and 3 of the
• the samples and the calibration standards use the same method called
For sample type B:
• there are 3 calibration standards of different concentrations and 9 samples,
• each sample must be analyzed once,
• the calibration standards must be analyzed twice and re-analyzed after
• the calibration standards of type B are in positions 4, 5 and 6 of the
• the samples and calibration standards of type B use different methods.
The samples use METH2 and the calibration standards use METH3. These
methods contain the same analytical parameters and differ only in the
analysis stoptime.
1200 Series Handheld Control Module Users's Guide
after every 2 samples,
C1 C2 C3 S7 S8 C1 C2 C3 S9 S10 C1 C2 C3 S11 S12 C1 C2 C3
autosampler tray and the 6 sample vials are in positions 7 to 12, and
METH1.
every 3 samples,
C1 C2 C3 S13-S15 C1 C2 C3 S16-S18 C1 C2 C3 S19-S21 C1 C2 C3
autosampler tray and the 9 sample vials are in positions 13 to 21, and
Running Multiple-Vial Analyses
Multiple-Level Calibration Sequences