The channels of a pumping system are named A, B, C and D (depending on the pump type).
N O T E
%A is automatically calculated by 100% - (%B + %C + %D). If no values for %B, %C and %D
are entered, %A is always 100%.
13 Select the Pressure button.
14 Set the upper pressure limit to 400 bar.
15 Select the Done button.
16 Select the Timetable button.
17 Select the Insert button.
18 Enter a value of 2 minutes and select Composition as the setting.
19 Select the Enter button to accept entries (Action Key F7).
20 Enter the following information for another two lines.
21 Press the Done button and verify your timetable entries.
22 Press the Esc key until you are in the Analysis screen.
Use Enter to store timetable lines.
N O T E
Use Done when all lines have been entered.
By pressing the Esc key you return to the previous screen.
1200 Series Handheld Control Module Users's Guide
If you have a quaternary pump set %B to 65, %C and %D to OFF.
If you have a binary pump set %B to 65.
If you have a quaternary pump, set the %B composition to 65, %C and %D to
0.00.
If you have a binary pump, set the %B composition to 65.
time 4 : %B = 95
time 5 : %B = 65
Running a Gradient Analysis
Entering Settings
10
127