Wait for Repeated Start Request
After pressing the Start button [F8] the autosampler will wait for external
start requests before each vial in the range or sequence. They have to be
generated by the external device. This mode is necessary when the external
device needs extra time to get ready for the next analysis and thus has to be in
charge of the start event.
Figure 21
Wait for repeated external start request
An Agilent 1200 variable wavelength detector or diode array detector will perform a balance
N O T E
(provided that Auto Balance is set to Prerun in the More ... Settings screen) when receiving
a start command from the control module. This will only happen in the Standard and Wait
for single (repeated) start request modes. In the Send single (repeated) start request modes
a balance before the run will NOT be performed. If regular balancing is required, set the
Auto Balance check box to Postrun.
1200 Series Handheld Control Module Users's Guide
Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices