Restrictions - Agilent Technologies G1323B User Manual

1200 series handheld control module
Table of Contents

Advertisement

2
Working with the Control Module
Coexecution with Agilent ChemStation

Restrictions

• If a parameter window is open for parameter entry on the Agilent
• If an analysis is running with the control module, the Agilent ChemStation
• If the Agilent ChemStation starts an analysis, the control module is the
• Parameter changes to a method will be identified on the other user
• The control module and the Agilent ChemStation have a different method
A method that is available on the control module as protected method can be
WA R N I N G
modified by the Agilent ChemStation and then be saved on the control module
without any warning.
62
ChemStation, this specific entry field is disabled on the control module.
should not be turned on or rebooted.
slave/monitor system.
interface as modification.
handling (Agilent ChemStation method can have more information than the
method on the control module, e.g. additional DAD parameters that are only
accessible from the Agilent ChemStation). To have a method available on
both controllers proceed as follows:
If the method is on the Agilent ChemStation and should be saved on the
control module or PC card, load the method on Agilent ChemStation and
then save the method on the control module (or PC card) with Method —
Save As.
If the method is on the control module or PC card and should be saved on
the Agilent ChemStation, first load method DEF_LC.M on the Agilent
ChemStation (to have no additional parameter in the format) and then load
the required method on control module. Then save the method on the
Agilent ChemStation with the same name.
1200 Series Handheld Control Module Users's Guide