Agilent Technologies G1323B User Manual page 140

1200 series handheld control module
Table of Contents

Advertisement

11
Running Multiple-Vial Analyses
Multiple-Level Calibration Sequences
3 Select the Calibration button [F1] to display the Calibration Settings screen for
4 Enter the following information:
5 Select the Done button [F6] to accept entries.
6 Select the Start button [F8] in the Sequence screen.
7 Press Enter start the
The autosampler now analyzes:
• the three calibration standards in duplicate,
• sample vials 10 through 14,
• the three calibration standards in duplicate,
• sample vials 15 through 19,
• the three calibration standards in duplicate,
• sample vials 20 through 24, and
• the three calibration standards in duplicate.
140
sequence line 1 and choose Edit.
Calibration vial range 90 to 92
Number of injections 2
Calibration method
METH1
Recalibrate every
5 vials multi
Before
On
After
Off
sequence.c
1200 Series Handheld Control Module Users's Guide