Multiple-Level Calibration Sequences

The following sections describe how to set up calibration sequences for
analyses which use multiple-level calibration.

Recalibrating With the Same Group of Standards

There are three calibration standards of different concentrations (C1, C2, C3)
and 15 samples (S). The standards and samples are analyzed using the same
method.
The analysis requires that:
• each sample is analyzed once,
• the calibration standards are analyzed twice before the samples and
• the calibration standards are located in positions 90 to 92 of the
• the 15 sample vials are located in positions 10 to 24 of the autosampler tray,
• the samples and standard are analyzed using a method called METH1.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence button [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
1200 Series Handheld Control Module Users's Guide
re-analyzed twice after every 5 samples,
C1 C2 C3 S10-S14 C1 C2 C3 S15-S19 C1 C2 C3 S20-S24 C1 C2 C3
autosampler tray,
and
Vial Range
10 to 24
#Inj.
2
Inj. Volume
Default
Method
METH1
Running Multiple-Vial Analyses

Multiple-Level Calibration Sequences

11
139