11
Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices
Send Repeated Start Request
This will cause the control module to generate start requests before each run.
The external device starts each injection then by sending a start signal to the
APG remote line. That is, after all the programming has been completed on the
control module and the Start button [F8] has been pressed, a Start Request is
issued before each run and the external device must give the Start command
for the injection process.
Figure 19
Send repeated external start request
Here, too, the module will simply track the progress of the analysis.
Wait for Single (External) Start Request
After pressing the Start button [F8] the autosampler will wait for a single
external start request on the APG remote lines. When the start request is
received, the complete vial range or sequence is done as in standard mode
under the command of the control module.
Figure 20
Wait for single external start request
146
1200 Series Handheld Control Module Users's Guide