Making An Injection

The sample loop can be either completely or partially filled. This depends on
the amount of sample you have.

Completely Filling the Sample Loop

This is the conventional method in which an excess of sample is used to fill the
sample loop completely. The volume of the loop determines the injection
volume.
1 Fill syringe with sample.
2 Move valve to LOAD position.
3 Insert needle of syringe into needle port until needle touches stator face.
4 Slowly inject sample.
5 Leave syringe in position and move valve to INJECT position.
6 Remove syringe.
1200 Series Handheld Control Module Users's Guide
Do not press too hard.
Using the Manual Injection Valve

Making An Injection

6