Tracking Autosampler History / EMF limits

1 Select the Records button [F4] in the System screen.
2 Select Autosampler from the list.
3 Press Enter.
You can check the number of 'Needle into Seat' injections the autosampler has
made and the number of injection valve cycles. A limit can be specified for
each function which you can use to schedule preventive maintenance.

Resetting the Autosampler

1 Select the Control button [F1] in the System screen.
2 Select Autosampler from the menu.
3 Press Enter.
4 Select the Reset button [F7] to reset the autosampler hardware.
This resets the injection valve, metering device and sampling unit.

Reset to Default Settings

1 Select the Settings button [F1] in the Analysis screen.
2 Select Autosampler from the menu.
3 Press Enter.
4 Select the Default button [F7] to reset the autosampler settings to their
1200 Series Handheld Control Module Users's Guide
default values.