Configuring the Analog Channel Output
You may use the LC Series 1200 detectors to test or monitor optical
parameters on external analog equipment connected via the analog output
(Fluorescence Light, Diode Array and Multiple Wavelength Detectors have two
analog outputs, Variable Wavelength Detector and Refractive Index Detector
have one) at the back side of the module.
To configure the analog outputs:
1 Press the Settings button [F1] in the Analysis screen.
2 Choose the detector and press enter.
3 Press the Analog button [F3].
4 Either select the signal you want to monitor from the signal source drop
5 Enter the values for Zero Offset (sets a baseline offset in order to recognize
6 Press the Done button [F6].
1200 Series Handheld Control Module Users's Guide
down list. Or, if you have a variable wavelength detector (fluorescence light
detector), choose the signal(s) from the list(s) to be routed to the analog
output(s).
negative drifts) and Attenuation (sets an absorbance range) for each of the
two connectors.
Using the Detectors
Configuring the Analog Channel Output
7