Installation

Installing the Calibration Factors

Installing the Calibration Factors
Cal factors can be entered and stored in an internal table (Automatic
Cal Factors) in the PSA's non-volatile memory. The table is
automatically accessed during RF power measurements (after the
measuring receiver has tuned to the frequency) to correct for the
non-linear response of the power sensor. If no entry in the table directly
corresponds to the frequency being measured, the measuring receiver
derives a cal factor through linear interpolation using the nearest
calibration points in the table.

Installing the Cal Factors in the N5530S

There are two methods of installing the sensor module calibration
factors (cal factors) in the N5530S Measuring Receiver System.
The first method loads and saves the cal factors using the PC. The
second method uses the PSA to load the data.
Figure 2-1
Utilities Menu
Both methods are accessed through the Utilities drop-down menu (See
Figure 2-1, "Utilities
Using the PC to Load and Save the Cal Factors
1. Insert the N5532A Sensor Module Data Disk in the PC A:\ drive.
NOTE
If your PC does not have a floppy disc drive, use the PSA to save your
cal data to your PC. Then use this procedure to load the sensor cal
factors into the system.
2. Select Calibration Factors to bring up the Calibration View
Edit dialog box.
3. Press Load Cal File in the Calibration View Edit dialog box. The
Open Calibration File dialog box appears.
4. Navigate to the PC A:\ drive if necessary and select the CFData.XML
file.
12
Menu.".)
Chapter 2