Setting up the Monitor
18
Caution: Use only the hospital-grade power cord provided by
Nellcor.
1. Power Connector
1. Set the SUPPLY VOLTAGE SELECTOR (2) switch to the
applicable voltage.
2. Plug the female connector end of the power cord into the N-595
POWER CONNECTOR (1) on the rear of the monitor.
3. Plug the male connector of the power cord into a properly
grounded AC outlet.
4. Verify that the monitor's AC POWER INDICATOR is lit.
Note: If the AC POWER INDICATOR is not lit, check:
the power cord
the SUPPLY VOLTAGE SELECTOR switch
the user-accessible fuses
the AC power outlet
2. Supply Voltage Selector