Description of Controls, Indicators, and Symbols
motion indicator and the pulse search indicator are simultaneously lit,
it is an indication that the artifact is significant and/or has been
persistent.
The Pulse Search Indicator. Lights continuously prior to initial
acquisition of a pulse signal and during prolonged and challenging
monitoring conditions. The pulse search indicator flashes during a
loss-of-pulse signal.
The Data In-Sensor Indicator. Lights to indicate that the attached
O
M
sensor contains a patient sensor event record. The sensor
XI
AX
event record information may be viewed or printed.
The SatSeconds Indicator. Fills in clockwise as the SatSeconds alarm
management system detects a %SpO
reading outside of the limit
2
setting. Empties in counterclockwise direction when %SpO
reading
2
is within limits. When the indicator is full, a medium priority alarm
will sound.
The Fast Response Mode Indicator. The response mode setting
dictates the response time (2 to 4 seconds in fast mode and 4 to 7
seconds in normal mode) applied by the O
M
algorithm in its
XI
AX
calculation of SpO
. The O
M
algorithm's calculation of pulse rate
2
XI
AX
is unaffected by the response mode setting. The trending interval
(2-seconds or 4-seconds) is updated automatically by the monitor to
roughly correspond with the SpO
calculation response time.
2
The Neonate Alarm Limits Indicator. This symbol is displayed when
the alarm limits are set to neonate. No symbol is displayed when the
monitor is set to adult limits.
1
Artifacts are events contained in the in-sensor data.