Table 14: Electromagnetic Immunity - Nellcor OxiMax N-595 Operator's Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Specifications
156

Table 14: Electromagnetic Immunity

-
The N
595 is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
is used in an electromagnetic environment as described below.
IEC
Immunity
60601
Test
Test Level
Electrostatic
±6 kV
discharge
contact
(ESD)
±8 kV air
IEC
-
-
61000
4
2
Electric fast
±2 kV for
transient/
power supply
burst
lines
IEC
±1 kV for
-
-
input/output
61000
4
4
lines
Surge
±1 kV
differential
mode
IEC
-
-
61000
4
5
±2 kV
common
mode
Note: UT is the AC mains voltage prior to application of the test level.
Compliance
-
-
1
2
Level
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±1 kV
differential
mode
±2 kV
common mode
-
595 should assure that it
Electromagnetic
Environment
Guidance
Floor should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least 30
%.
Mains power quality
should be that of a
typical commercial
and/or hospital
environment
Mains power quality
should be that of a
typical commercial
and/or hospital
environment

Hide quick links: