Nellcor OxiMax N-595 Operator's Manual page 131

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Table 10: Prompt/Error Messages
Message
DEFAULTS
LOST
DELETE
TRENDS?
LOW
BATTERY
(1) The first press of the Alarm Silence softkey will silence any audible tone and the second
press will clear the message.
READING
TRENDS ...
SENSOR
DISCONNECT
ED
N-595
-
Time
Exit on
Exit on
Alarm
out
Alarm
Silence
(sec.)
None
No
No
10
Yes
Yes
None
No
Yes (1)
None
Yes
Yes
None
No
Yes
When Displayed
-
If the N
595
detects that
-
power
on settings
have been lost.
When the user
attempts to delete
trend data from
memory by
pressing the
DELETE softkey.
When the monitor
is on battery power
and the battery
charge is low.
-
When the N
595
needs more than 4
to 6 seconds to
retrieve trend data
from memory.
1
When the sensor is
disconnected from
the monitor.
Troubleshooting
How Cleared
After the monitor
-
is power
cycled.
After the user
responds to the
prompt.
When the
monitor is
connected to AC
power or when
the low battery is
acknowledged
by pressing the
ALARM
SILENCE
button.
When sensor
data is
completely
retrieved or
ABORT is
selected.
When the sensor
is reconnected or
when the sensor
disconnection is
acknowledged
by pressing the
ALARM
SILENCE
button.
121