4. If the N-595 detects an internal problem during POST, an error
tone sounds and the monitor displays an error code (EEE) and the
corresponding number (see Troubleshooting on page 117).
5. Upon successful completion of the POST, the N-595 sounds a
one-second tone indicating that the monitor has passed the test.
WARNING: If you do not hear the POST pass tone, do not use
the pulse oximeter.
WARNING: Ensure that the speaker is clear of any obstructions.
Failure to do so could result in an inaudible alarm tone.
Note: In addition to serving as the POST pass verification, the POST
pass tone also functions as an audible confirmation that the
speaker is performing properly. If the speaker does not
function, the alarm warning sounds cannot be heard.
When an O
M
sensor is attached to the monitor, a "DATA TYPE: .
XI
AX
. ." message is displayed briefly at the bottom of the monitor display.
For a sensor containing data, the message identifies the sensor data
type, For a blank sensor, the message identifies the monitor's current
Using the Monitor
31