Real-Time Display/Printout Format

N-595
The graph mode disables all printout functions except trend data.
Graph mode trend printouts are formatted for a Seiko DPU-414 and
Okidata 320 serial printer.

Figure 4: Graph Mode Printout

Real-time data is continuously sent to the data port on the back of the
N-595. Patient data can be obtained through the data port by
connecting the monitor data port to a PC or serial printer. When a
real-time printout or display is being transmitted to a printer or PC, a
new line of data is displayed every 2 seconds. Column headings are
displayed or printed after every 25 lines, or if one of the values in the
column heading changes. Readings are displayed at 4-second
intervals if the SpO
response mode is set to normal and at 2-second
2
intervals when the SpO2 response mode is set to fast.
Data cannot be obtained if the N-595 is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the monitor is connected to a PC, send an
XON (Ctrl-q) command to reset the monitor.
Printing
83