LIMITS
ADULT or
NEO
SELECT
EXIT
N-595
1. Press the LIMITS softkey. Current alarm limits are displayed.
or
2. Press the ADULT or NEO softkey to select Adult-Pediatric or
Neonatal alarm limits screen.
3. Press the SELECT softkey as required to select the parameter to
be adjusted.
4. Use the ADJUST UP or ADJUST DOWN buttons to increase or
decrease the selected limit parameter.
5. Repeat steps 3, 4, and 5 as necessary to complete the alarm limits
setup.
6. To accept the changes, let the display time-out or press the EXIT
softkey to exit the display and return to normal monitoring.
Using the Monitor
45