N-595
Table 14: Electromagnetic Immunity
-
The N
595 is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
is used in an electromagnetic environment as described below.
IEC
Immunity
60601
Test
Test Level
Voltage dips,
<5 % U
short
interruptions
and voltage
variations on
(>95 % dip in
power supply
U
) for 0.5
T
cycle
40 % U
IEC
-
-
61000
4
11
(60 % dip in
U
) for 5
T
cycles
70 % U
(30 % dip in
U
) for 25
T
cycles
<5 % U
(95 % dip in
U
) for 5 sec.
T
Note: UT is the AC mains voltage prior to application of the test level.
Compliance
-
-
1
2
Level
<5 % U
T
T
(>95 % dip in
U
for 0.5
T)
cycle
40 % U
T
T
(60 % dip in
U
for 5
T)
cycles
70 % U
T
T
(30 % dip in
U
for 25
T)
cycles
<5 % U
T
T
(95 % dip in
U
for 5 sec.
T)
Specifications
-
595 should assure that it
Electromagnetic
Environment
Guidance
Mains power quality
should be that of a
typical commercial
and/or hospital
environment. If the
-
user of the N
595
requires continued
operation during
power mains
interruption, it is
recommended that
-
the N
595 be
powered from an
uninterruptible
power supply or
battery.