Table 9: Error Codes - Nellcor OxiMax N-595 Operator's Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Troubleshooting
118
Table 9 on page 118 lists the error codes and possible causes. When
this occurs, the unit will stop monitoring, remove all information
from the screen and display the message "EEE XXX," and sound a
low priority alarm. Cycling the power clears these errors.

Table 9: Error Codes

Error
Error
Code
Message
80
DEFAULTS
LOST
81
SETTINGS
LOST
82
CLOCK
SETTING
LOST
530
LOW
BATTERY
Action
-
The current power
on default settings
have been lost and returned to factory
defaults. Qualified service personnel
can use the service manual to restore the
-
desired power
on default settings.
The current settings (for example, alarm
limits, alarm and pulse beep volumes,
alarm silence duration) have been lost
-
and returned to power
on defaults. Turn
the monitor off and back on again. If it
is necessary to have settings different
-
from the power
on default settings, turn
the monitor off and back on again, and
reenter the desired settings.
The date and time settings have been
lost. Reenter the date and time.
The battery is discharged to a critically
low level.
Check to ensure that the voltage
selector switch on the rear panel is set to
the proper voltage.
Turn the monitor off and let it charge
for approximately 10 minutes and then
reattempt to turn the unit on. If the error
code is still present, turn the unit off and
let it continue to charge. If the monitor
has been charged for 30 minutes and the
error code is still present, notify service
personnel.