Trigger
Trigger
Trigger Mode
Note: When doing calibration or read rate testing, the current trigger setting will be
disregarded.
Note: This command does not need to be followed by an <A> command.
Definition:
The type of trigger event that will initiate the read cycle.
Serial Cmd:
<K200,trigger
Default:
Continuous Read
Options:
0 = Continuous Read
1 = Continuous Read 1 Output
2 = External Level
Continuous Read
Usage:
Useful in testing bar code symbol readability or scanner functions. It is
not recommended for normal operations.
Definition:
In Continuous Read, trigger input options are disabled, the scanner is
always in the read cycle, and it will attempt to decode and transmit
every scan crossing a symbol.
When To Output and Noread options have no affect on Continuous
Read.
Serial Cmd:
<K200,
4-6
MS-9 Scanner User's Manual
mode,trigger filter duration>
0>
3 = External Edge
4 = Serial Data
5 = Serial Data & Edge