Pulsed IV System Setup

Initially, the following system setup is required to use the pulsed IV test system. For
the timing of the system setup, see Table 3-1.
System Configuration
This action is necessary to specify the GPIB address and the source or
measurement channels of PGU (pulse generator), DSO (digital sampling
oscilloscope), and SMU (source monitor unit) used for the pulsed IV
measurement. See "System Configuration" on page 3-10.
Compensation
This action is necessary to obtain and update the compensation factors of the
pulsed IV test system. See "Compensation" on page 3-11.
Drain Cable Replacement
This action is necessary to obtain and update the compensation factors of the
bias-T and the cable of the drain measurement path. See "Drain Cable
Replacement" on page 3-16.
Execution Timing and Required Action
When the system setup is required
after system installation
after changing GPIB address of an instrument
after changing a channel assignment
after repair of oscilloscope
after calibration of oscilloscope
after replacing oscilloscope
after replacing bias-T
after replacing measurement cables to drain
Agilent B1542A User's Guide, Edition 4
Performing System Setup and Compensation
"System Configuration"
"Compensation"
"System Configuration"
"System Configuration"
"Compensation"
"Compensation"
"Compensation"
"Compensation"
"Drain Cable Replacement"
"Drain Cable Replacement"
Pulsed IV System Setup