Drain Cable Replacement - Agilent Technologies B1542A User Manual

Agilent b1542a pulsed iv package for b1500a/easyexpert
Table of Contents

Advertisement

Performing System Setup and Compensation

Drain Cable Replacement

Drain Cable Replacement
Perform this action after replacing the bias-T or cables used for the drain terminal
measurement path.
1. Open the PLSDIV Setup test setup from your preset group.
2. Set the Action field to Drain Cable Replacement.
3. Click the Single button at the upper right corner of the EasyEXPERT/Desktop
EasyEXPERT screen. The connection diagram window is displayed. See Figure
3-6 for the pulsed IV test system or Figure 3-7 for the DC + pulsed IV test
system.
3- 16
Agilent B1542A User's Guide, Edition 4