Execution
Conditions
Filename
Write Type
Read Type
Write String
Read List
Response

PlsdivCaptureMonData

This command reads the oscilloscope vertical data (monitor data). It returns the gate
voltage data, drain voltage data, or drain current data selected by the term and
drain_mode parameters.
PlsdivSetSampling and PlsdivMeasureId must be executed before this command.
The PlsdivSetSampling command sampling_num parameter value must not be 0.
PlsdivCaptureMonData.exe
String
List
This command needs the following input parameters in this sequence.
term
Terminal. Numeric value, 0 (gate) or 1 (drain).
sampling_num
Sampling number of oscilloscope. Numeric value, 16 to
262144. The value must be the same as the sampling_num
parameter value set to the PlsdivSetSampling command.
drain_mode
Drain data type. Numeric value, 0 (voltage) or 1 (current).
If term=0, this command returns the gate voltage data.
If term=1 and drain_mode=0, this command returns the drain voltage data.
If term=1 and drain_mode=1, this command returns the drain current data.
This command returns the following output parameter.
read_data
Oscilloscope vertical data (monitor data). Vector data. The
number of returned data is sampling_num if the smoothing is
off or the number of data displayed on the oscilloscope screen if
the smoothing is on. The smoothing on/off is set by the
PlsdivSetSmoothing command.
Number of returned data. Numeric value.
Agilent B1542A User's Guide, Edition 4
PLSDIV TIS Commands
PlsdivCaptureMonData
6- 11