PLSDIV TIS Commands

PlsdivForceVd

PlsdivForceVd
This command sets the SMU (source monitor unit) output switch on and applies the
specified drain voltage.
PlsdivForceVd.exe
String
String. However, this command does not return any output parameter.
This command needs the following input parameter.
vd
Drain voltage in V. Numeric value, -10 to 10.
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.

PlsdivForceVg

This command sets the PGU (pulse generator unit) output switch on, forces the gate
voltage pulse, and returns the gate voltage monitor value.
PlsdivSetPulseWidth and PlsdivSetVgLevel commands must be executed before
this command in this sequence.
PlsdivForceVg.exe
String. However, this command does not need any input parameter.
List
This command returns the following output parameters in this sequence. To read the
values, specify a vector variable in the Values field of the ReadList area. The values
are stored in the first and second elements of the vector variable.
vg_mon
Gate voltage monitor value that is the actual gate voltage
monitored by the oscilloscope. Numeric value, in V.
conv_stat
Convergence result of the gate voltage automatic adjustment.
Numeric value, 0 (convergence condition is satisfied) or -1 (not
satisfied). If the automatic adjustment is off, 0 is returned. The
convergence condition is a system parameter, no user definable.