Auto Adjustment
Response
Filename
Write Type
Read Type
Write String
Gate Zero
Connection Check This function is used to confirm if the divider, terminator, and bias-T are connected
This function is set by the PlsdivSetAdjustStat command, and is used to adjust the
gate voltage automatically so that the vg_mon value is close to the vg_high value
specified by the PlsdivSetVgLevel command.
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.

PlsdivInit

This command initializes the pulsed IV test system. This command must be
executed before any other Plsdiv command.
PlsdivInit.exe
String
String. However, this command does not return any output parameter.
This command needs the following input parameters in this sequence.
connect_check
Connection check status. Numeric value, 0 (does not perform)
or 1 (performs connection check in the PlsdivInit process).
gate_zero
Gate zero status. Numeric value, 0 (does not perform) or
1 (performs gate monitor channel offset measurement in the
PlsdivInit process).
This function is used to measure the offset data of the gate voltage monitor path and
update the offset value in a system file. The offset value is used until the next
execution. Before performing the offset measurement, set up the test system
measurement path properly and open the measurement terminals.
properly. For the proper connections, see Figure 6-2 and Figure 6-3. The connection
check can detect the following improper connections.
divider is not connected
terminator is not connected
T-adapter is connected instead of divider
bias-T's DC+RF connector is connected to the gate monitor channel
Agilent B1542A User's Guide, Edition 4
PLSDIV TIS Commands
PlsdivInit
6- 15