Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.

PlsdivSetDeviceDelay

This command sets the device delay time between the gate pulse and the drain pulse.
Before performing the pulsed IV measurement, figure out the device delay time and
set it by using this command. To figure out the device delay time, set up the test
system properly, connect the device under test, apply voltage pulse to the gate
terminal, and monitor the gate pulse and the drain pulse as shown in Figure 6-4.
PlsdivSetDeviceDelay.exe
String
String. However, this command does not return any output parameter.
This command needs the following input parameter.
device_delay
Delay time caused by test device, in second. Numeric value,
-0.000001 to 0.000001.
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Adjustment for Device Delay