To set up the system (System Setup)

3200P pH Meter User Guide
System setup includes the following items:
Calib Interval
Calib Interval is the time interval between
recalibration prompts. The meter counts
the time since the last calibration and if
the time counted exceeds the entered
calibration time interval, the meter
displays a popup window to remind the
user to recalibrate the electrode. Calib
Interval is measured in hours. Input zero
(0) hours to disable.
Operator No
Operator No is a three- digit number in the
range 000 to 200. Every set of recorded
data contains the Operator No. field. This
field is used to record the operator
performing measurements for regulated
(GLP) laboratories. If you need to track
the operator, assign a unique number for
each operator.
Time
System time and date. See
and time"
Auto Delete
Enable to automatically overwrite the
oldest stored data entries with new data if
the data storage is full.
1 Press [Setup].
2 Select System Setup, and press [Enter] to enter System Setup.
See
Figure
14.
"To set the date
on page 30.