Connecting Switch Controller (For B1542A-024) - Agilent Technologies B1542A User Manual

Agilent b1542a pulsed iv package for b1500a/easyexpert
Table of Contents

Advertisement

Figure 2-11
Installation
Hardware Installation

Connecting Switch Controller (for B1542A-024)

1. Install the Agilent 11713A switch controller and connect the GPIB cable. See
"Installing Instruments" and "Connecting GPIB Cables" on page 2-6.
2. Fix the switch control distributor to the appropriate place.
3. Connect the viking cable between the 11713A switch controller and the switch
distributor.
4. Connect the D-sub 9 pin cable between the switch distributor Switch 1
connector and the gate pulse/dc switch unit.
5. Connect the D-sub 9 pin cable between the switch distributor Switch 2
connector and the drain pulse/dc switch unit.
Required accessories and equipment:
Viking cable, 1 ea.
D-sub 9 pin cable, 2 ea.
Agilent 11713A Switch Controller, 1 ea.
Switch Control Distributor, 1 ea.
Gate Pulse/dc Switch Unit with divider, 1 ea.
Drain Pulse/dc Switch Unit with bias-T, 1 ea.
Switch Controller, Switch Control Distributor, and Pulse/dc Switch Units
2- 14
Agilent B1542A User's Guide, Edition 4