Figure 6-7
Response
Filename
Response
PLSDIV TIS Commands
PlsdivUninit
vg_base
Pulse base level, in V. Numeric value, -4.5 to 4.5.
polarity
Polarity of pulse. Numeric value, 0 (positive) or 1 (negative).
See Figure 6-7.
where, vg_high > vg_base and 55 mV < |vg_high - vg_base| < 4.5 V must be
satisfied.
To use the gate voltage automatic adjustment function, set |vg_high - vg_base| <
4.0 V.
Pulse Level and Polarity
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivUninit
This command resets the pulsed IV test system (all instruments) and closes GPIB
communication to each instrument.
After executing this command, any Plsdiv commands except for PlsdivInit cannot
be executed. So, make sure to execute PlsdivInit before executing any command.
PlsdivUninit.exe
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
6- 24
Agilent B1542A User's Guide, Edition 4