System Configuration - Agilent Technologies B1542A User Manual

Agilent b1542a pulsed iv package for b1500a/easyexpert
Table of Contents

Advertisement

Table 3-2
Field name
Controller
Visa Name
Instrumentations
a
GPIB Address
CH Assignment
a. To know the instrument GPIB address, see the manual of each instrument.
Performing System Setup and Compensation

System Configuration

System Configuration
Perform this action after completing the system installation or after changing a
channel assignment or after changing the GPIB address of the system components.
1. Open the PLSDIV Setup test setup from your preset group.
2. Set the Action field to System Configuration.
3. Click the Single button at the upper right corner on the EasyEXPERT/Desktop
EasyEXPERT screen. The Configuration dialog box is displayed.
4. Set the system configuration parameters by using the Configuration dialog box.
See Table 3-2.
System Configuration Parameters
PGU
DSO
SMU
USB Address
PGU, DSO, and SMU
Gate PGU CH
Gate Monitor DSO CH
Drain Monitor DSO CH
Drain Bias SMU CH
3- 10
Description
Select B1500A Internal or External PC.
Enter the VISA interface ID of the GPIB interface.
Specify the pulse generator to be used.
Specify the oscilloscope to be used.
Specify the DC source monitor to be used.
If Controller=B1500 Internal, enter the B1500A's
internal USB address.
Enter GPIB address of pulse generator (PGU),
oscilloscope (DSO), and DC source monitor (SMU).
Enter DSO channel number for Gate pulse output.
Enter DSO channel number for Gate voltage monitor.
Enter DSO channel number for Drain voltage monitor.
Enter SMU slot number for Drain voltage output.
Agilent B1542A User's Guide, Edition 4