Figure 5-1
Device Parameters The following parameters are available to set the test condition.
PLSDIV Test Definitions

PLSDIV Capt Wave

PLSDIV Capt Wave
This test definition controls SMU (source monitor unit) to force the drain voltage,
controls PGU (pulse generator) to force the gate voltage pulse, controls the
oscilloscope to monitor the gate voltage pulse waveform and the drain voltage pulse
waveform, calculates the drain current pulse waveform if the drain monitor mode is
current, and displays the gate voltage pulse waveform and the drain voltage or
current pulse waveform.
Pulse Waveform Measurement Result Example
Polarity
Polarity of source output. Nch (force the specified value) or Pch
(force the negative specified value). See "Parameters" on page
5-31.
Gate length, in m. ≥ 1 nm, 4 digits or 1 nm resolution.
Lg
Gate width, in m. ≥ 1 nm, 4 digits or 1 nm resolution.
Wg
Temp
Temperature, in degree. 4 digits or 0.001 degree resolution.
IdMax
Maximum drain current measurement range, in A. 0 to 80 mA,
4 digits or 1 μA resolution.
Delay time caused by device under test, in s. -1 μs to 1 μs,
DeviceDelay
4 digits or 100 ps resolution. See "Parameters" on page 5-31.
5- 4
Agilent B1542A User's Guide, Edition 4