Figure 2-7

Installing Pulsed IV Test System

The connection overview of the pulsed IV test system (without the option B1542A-
024) is shown in Figure 2-5. To install the system, perform the following steps.
1. "Making Measurement Path for Gate" on page 2-9
2. "Making Measurement Path for Drain" on page 2-10
3. "Connecting RF Probes" on page 2-17 or "Connecting DC Probes" on page 2-18
Making Measurement Path for Gate
Connect the following accessories as shown in Figure 2-7. And fix the divider to the
appropriate place. Use a torque wrench and an open-end wrench to fasten the SMA
connectors.
Required accessories:
BNC(m)-SMA(f) adapter, 1 ea., for pulse generator
BNC(m)-SMA(f) adapter, precision type, 1 ea., for oscilloscope
BNC cable, 1 ea.
SMA cable, 0.3 m, 1 ea.
SMA cable, 1.5 m, 1 ea.
Divider, 1 ea.
Measurement Path for Gate
Agilent B1542A User's Guide, Edition 4
Installation
Hardware Installation
2- 9