Figure 2-12
DC source monitor
(*1) Use BNC(m)-SMA(f) adapter
for pulse generator.
(*2) Use BNC(m)-SMA(f) adapter
precision type for oscilloscope.
Making Measurement Path for Gate (for B1542A-024)
Connect the following accessories as shown in Figure 2-12. And fix the gate
pulse/dc switch unit to the appropriate place. Use a torque wrench and an open-end
wrench to fasten the SMA connectors.
Required accessories:
•
BNC(m)-SMA(f) adapter, 1 ea., for pulse generator
•
BNC(m)-SMA(f) adapter, precision type, 1 ea., for oscilloscope
•
BNC cable, 1 ea.
•
SMA cable, 0.3 m, 1 ea.
•
SMA cable, 1.5 m, 1 ea.
•
Triaxial cable, 1 ea.
Measurement Path for Gate
to SMU Force
for gate voltage
at DC test
Agilent B1542A User's Guide, Edition 4
Pulse generator
Scope
to input channel
for gate (*2)
Triaxial cable
SMU
Hardware Installation
to output channel for gate pulse (*1)
to AUX Trig In
SMA cable, 1.5 m
SMA cable, 0.3 m
Gate Pulse Monitor
Gate pulse/dc switch unit
Gate Pulse Input
Installation