Transfer Accuracy - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

AC Accuracy (continued): 24 Hour to 2 Year (% of Reading + % of Range)
Range
45 Hz to 100 kHz 100 kHz to 1 MHz 1 MHz to 4 MHz 4 MHz to 8 MHz 8 MHz to 10 MHz
10 mV
0.09 + 0.06
100 mV – 10 V 0.09 + 0.06
100 V
0.12 + 0.002
1000 V
0.3 + 0.01

Transfer Accuracy

Range
% of Reading
100 mV – 100 V (0.002 + Resolution in %)
AC + DC Accuracy (ACDCV Function)
For ACDCV Accuracy apply the following additional error to the ACV accuracy. (% of Range)
DC <10% of AC Voltage
ACBAND £ 2 MHz ACBAND > 2 MHz Temperature Coefficient
Range
10 mV
0.09
100 mV – 1000 V
0.008
DC >10% of AC Voltage
ACBAND £ 2 MHz ACBAND > 2 MHz Temperature Coefficient
Range
10 mV
0.7
100 mV – 1000 V
0.07
Additional Errors
Apply the following additional errors as appropriate to your particular measurement setup. (% of Reading)
Source R
0–1 MHz 1-4 MHz 4–8 MHz 8–10 MHz
0 W
0
50 WTerminated 0.003
75 WTerminated 0.004
50 W
0.005
4
Reading Rates
ACBAND Low
Maximum Sec / Reading
1 – 5 Hz
6.5
5 – 20 Hz
2.0
20 – 100 Hz
1.2
100 – 500 Hz
0.32
>500 Hz
0.02
Settling Characteristics
There is no instrument settling required.
Common Mode Rejection
For 1 kW imbalance in LO lead, > 90 dB, DC to 60 Hz.
ACBAND >2 MHz
1.2 + 0.05
7 + 0.07
2.0 + 0.05
4 + 0.07
Conditions
Following 4 Hour warm-up
1
Within 10 min and ±0.5°C of the reference measurement
45 Hz to 20 kHz, sine wave input
Within ±10% of the reference voltage and frequency
0.09
0.09
0.7
0.7
3
Input Frequency
2
5
5
0
0
0
2
5
5
3
7
10
% Resolution
0.001 – 0.005
0.005 – 0.01
0.01 – 0.05
0.05 – 0.1
0.1 – 1
>1
20 + 0.08
4 + 0.08
15 + 0.1
0.03
0.0025
0.18
0.025
Crest Factor Resolution Multiplier
1–2
(Resolution in%) × 1
2–3
(Resolution in%) × 2
3–4
(Resolution in%) × 3
4–5
(Resolution in%) × 5
Maximum Sec / Reading
32
6.5
3.2
0.64
0.32
0.1
Appendix A Specifications
1.
Resolution in % is the value
of RES command or
parameter (reading
resolution as percentage of
measurement range).
2.
Additional error beyond
±1°C, but within ±5°C of
last ACAL. (% of Range)/
°C. For ACBAND >2 MHz,
use 10 mV range
2
temperature coefficient. Lo
to Guard switch on.
2
3.
Flatness error including
instrument loading.
1
4.
Reading time is the sum of
the Sec / Reading shown for
your configuration. The
tables will yield the slowest
reading rate for your
configuration. Actual
reading rates may be faster.
For DELAY– 1; ARANGE
OFF.
289