Agilent Technologies 3458A User Manual page 175

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

weights. The error conditions and their weights are:
Weighted
Value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
Power-on value = 32767 (all enabled).
Default value = 32767 (all enabled).
Remarks
When an error occurs, it sets the corresponding bit in the error register
regardless of whether or not it has been enabled by the EMASK command.
Disabling an error bit prevents it from setting the error bit in the status register
only, and thereby generating a service request.
Query Command The EMASK? query command returns the weighted sum of
all enabled error conditions (see example below).
Related Commands: AUXERR?, ERR?, ERRSTR?, RQS, STB?
Examples
OUTPUT 722;"EMASK 4
OUTPUT 722;"EMASK 248 "
OUTPUT 722;"EMASK 0
10 OUTPUT 722; "EMASK?"
20 ENTER 722;A
30 PRINT A
40 END
Bit
Number
Error Conditions
0
Hardware error (see AUXERR? for more information)
1
Calibration error
2
Trigger too fast error
3
Syntax error
4
Command not allowed from remote (ADDRESS command)
5
Undefined parameter received
6
Parameter out of range
7
Memory Error
8
Destructive overload detected
9
Out of calibration
10
Calibration required
11
Settings conflict (memory improperly configured for
sub-sampling)
12
Math error (divide by 0, integer overflow, etc.)
13
Subprogram error (calling a deleted sub, CONT with no
PAUSE, SUBEND or PAUSE only allowed in sub,
SCRATCH, DELSUB, CONT, not allowed in sub)
14
System error
"
!ENABLES THE TRIGGER TOO FAST ERROR
!ENABLES ERRORS 8, 16, 32, 64, AND 128
"
!DISABLES ALL ERRORS
!RETURNS EMASK VALUE
!ENTER RESPONSE
!PRINT VALUE
Chapter 6 Command Reference
EMASK
175

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents