TRIG
258
Chapter 6 Command Reference
multimeter is properly configured. Line 20 suspends measurements by setting the
trigger event to HOLD. Lines 30 and 40 configure for 30 DC voltage readings per
trigger event. Line 50 generates a single trigger causing the multimeter to make
thirty readings. After the readings are complete, the trigger event reverts to HOLD.
10 OUTPUT 722;"RESET"
20 OUTPUT 722;"TRIG HOLD"
30 OUTPUT 722;"DCV 10"
40 OUTPUT 722;"NRDGS 30,AUTO"
50 OUTPUT 722;"TRIG SGL"
60 END
!RETURN TO POWER-ON STATE
!SUSPEND READINGS
!DC VOLTAGE MEASUREMENTS,10V RANGE
!30 READINGS PER SAMPLE EVENT (AUTO)
!GENERATES A SINGLE TRIGGER