The Display - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

The Display

26 Chapter 2 Getting Started
Command
ACBAND 20, 2E6
AC bandwidth 20Hz - 2MHz
AZERO ON
Autozero enabled
DCV AUTO
DC voltage, autorange
DEFEAT OFF
Defeat disabled
DELAY -1
Default delay
DISP ON
Display Enabled
EMASK 32767
Enable all error conditions
END OFF
Disable GPIB EOI function
EXTOUT ICOMP, NEG
Input complete EXTOUT signal, negative pulse
FIXEDZ OFF
Disable fixed input resistance
FSOURCE ACV
Frequency and period source is AC voltage
INBUF OFF
Disable input buffer
LEVEL 0, AC
Level trigger at 0%, AC-coupled
LFILTER OFF
Level filter disabled
LFREQ 50 or 60
Measured line frequency rounded to 50 or 60Hz
LOCK OFF
Keyboard enabled
MATH OFF
Disable real-time math
MEM OFF
Disable reading memory (last memory operation = FIFO)
MFORMAT SREAL
Single real reading memory format
MMATH OFF
Disable post-process math
NDIG 7
Display 7.5 digits
NPLC 10
10 power line cycles of integration time
NRDGS 1, AUTO
1 reading per trigger, auto sample event
OCOMP OFF
Disable offset compensated resistance
OFORMAT ASCII
ASCII output format
QFORMAT NORM
Normal query format
RATIO OFF
Disable ratio measurements
RQS 0 (or 8)
0 disables status register conditions (if power-on
SRQ was on when power was removed, value = 8).
SETACV ANA
Analog AC voltage mode
SLOPE POS
Positive slope for level triggering
SSRC LEVEL, AUTO
Level sync source event, auto synchronous AC voltage
SWEEP lOOE-9,1024
Sample interval 100 nanoseconds, 1024 samples
TARM AUTO
Auto trigger arm event
TBUFF OFF
Disable external trigger buffering
TIMER 1
1 second timer interval
TRIG AUTO
Auto trigger event
All math registers set to 0 except:
DEGREE = 20
In the power-on state, the display is continuously updated with each new DC
voltage reading. Along the bottom of the display are a series of annunciators.
These annunciators alert you to a variety of conditions. For example, the
SMPL annunciator flashes whenever the multimeter has completed a
reading. Table 6 describes the meaning of each display annunciator.
Table 5. Power-On State
Description
REF=l
SCALE = 1 RES=50
PERC = 1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents