SCAL

Remarks
Example

SCAL

SCRATCH

Syntax
Remarks
Example

SECURE

Syntax
232
Chapter 6 Command Reference
Power-on name = none.
Default name = 0.
Whenever the multimeter's power is removed, the present state is stored in state
0. After a power failure, the multimeter can be configured to its previous state
by executing RSTATE 0.
If the NULL real-time math operation was enabled in a stored state, after
recalling the state, the first reading is placed in the OFFSET register (refer to
"NULL" in Chapter 4 for more information).
From the front panel, you can review the names of all stored states by pressing
the Recall State key and by using the up and down arrow keys. When you have
found the desired state, press the Enter key to recall that state.
Related Commands: MSIZE, PURGE, SCRATCH, SSTATE
OUTPUT 722; "RSTATE B2"
This is a calibration command. Refer to the 3458A Calibration Manual for details.
Clears all subprograms and stored states from memory.

SCRATCH

Individual subprograms can be cleared with the DELSUB command. Individual
states can be cleared with the PURGE command.
Related Commands: DELSUB, PURGE, RSTATE, SSTATE, SUB
OUTPUT 722;"SCRATCH"
Security Code. Allows the person responsible for calibration to enter a security
code to prevent accidental or unauthorized calibration or autocalibration (autocal).
(Refer to the ACAL command for details on autocal.)
SECURE old_code, new_code[,acal_secure]
old_code
This is the multimeter's previous security code. The multimeter is shipped from
!RECALLS STORED STATE NAMED B2
!CLEARS ALL SUBPROGRAMS AND STORED STATES