operation
Numeric
Parameter
Equiv.
FTHRM
8
NULL
9
PERC
10
PFAIL
11
RMS
12
SCALE
13
STAT
14
CTHRM2K
16
CTHRM10K
17
FTHRM2K
18
FTHRM10K
19
CRTD85
20
CRTD92
21
FRTD85
22
FRTD92
23
Power-on operation_a,operation_b = OFF,OFF.
Default operation_a,operation_b = OFF,OFF.
Description
Result=temperature (Fahrenheit) of a 5kW thermistor
(40653B). Function must be OHM or OHMF (10kW)
range or higher).
Result=reading-OFFSET register. The OFFSET
register is set to first reading—after that you can
change it.
Result =((reading - PERC register) / PERC register)
x 100.
Reading vs. MAX and MIN registers.
Result = squares reading, applies FILTER operation,
takes square root.
Result = (reading-OFFSET register) / SCALE
register.
Performs statistical calculations on the present set of
readings and stores results in these registers:
SDEV = standard deviation
MEAN = average of readings
NSAMP = number of readings
UPPER = largest reading
LOWER = smallest reading
Result=temperature (Celsius) of a 2kW thermistor
(40653A). Function must be OHM or OHMF.
Result=temperature (Celsius) of a 10kW thermistor
(40653C). Function must be OHM or OHMF.
Result=temperature (Fahrenheit) of a 2kW thermistor
(40653A). Function must be OHM or OHMF.
Result=temperature (Fahrenheit) of a 10kW
thermistor (40653C). Function must be OHM or
OHMF.
Result=temperature (Celsius) of 100W RTD with
alpha of 0.00385 (40654A or 40654B). Function
must be OHM or OHMF.
Result=tempersture (Celsius) of 100W RTD with
alpha of 0.003916. Function must be OHM or
OHMF.
Result=temperature (Fahrenheit) of 100W RTD with
alpha of 0.00385 (40654A or 40654B). Function
must be OHM or OHMF.
Result=temperature (Fahrenheit) of 100W RTD with
alpha of 0.003916. Function must be OHM or
OHMF.
Chapter 6 Command Reference
MMATH
201