Delsub; Diagnost; Disp - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

DELSUB

Syntax
name
Remarks
Example

DIAGNOST

DISP

Syntax
control
Delete Subprogram. Removes a single subprogram from memory.
DELSUB name
Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the SUB command for details.
Power-on name = none.
Default name = none; parameter required.
When a subprogram is deleted, the memory used to store it is freed and may be
used to store a new subprogram (see the SUB command).
To delete all subprograms at once, use the SCRATCH command.
Related Commands: COMPRESS, SCRATCH, SUB
OUTPUT 722;"DELSUB TEST12"
This is a service-related command. Refer to the 3458A Service Manual for details
Display. Enables or disables the multimeter's display, and may also be used to
send a message to the display or to clear the display.
DISP [control] [,message]
The control parameter choices are:
Numeric
control
Query
Parameter
Equiv.
OFF
0
ON
1
MSG
2
CLR
3
!DELETES SUBPROGRAM TEST12
Description
Displays message if included (if no message, dashes are
displayed); inactivates all annunciators except ERR;
readings are no longer displayed and the display is not
updated except to service front panel keystrokes and
query commands.
Normal (power-on mode) display operation
Displays message, annunciators activated
Clears the display
Chapter 6 Command Reference

DELSUB

171

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents