DELSUB
Syntax
name
Remarks
Example
DIAGNOST
DISP
Syntax
control
Delete Subprogram. Removes a single subprogram from memory.
DELSUB name
Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the SUB command for details.
Power-on name = none.
Default name = none; parameter required.
•
When a subprogram is deleted, the memory used to store it is freed and may be
used to store a new subprogram (see the SUB command).
•
To delete all subprograms at once, use the SCRATCH command.
•
Related Commands: COMPRESS, SCRATCH, SUB
OUTPUT 722;"DELSUB TEST12"
This is a service-related command. Refer to the 3458A Service Manual for details
Display. Enables or disables the multimeter's display, and may also be used to
send a message to the display or to clear the display.
DISP [control] [,message]
The control parameter choices are:
Numeric
control
Query
Parameter
Equiv.
OFF
0
ON
1
MSG
2
CLR
3
!DELETES SUBPROGRAM TEST12
Description
Displays message if included (if no message, dashes are
displayed); inactivates all annunciators except ERR;
readings are no longer displayed and the display is not
updated except to service front panel keystrokes and
query commands.
Normal (power-on mode) display operation
Displays message, annunciators activated
Clears the display
Chapter 6 Command Reference
DELSUB