Agilent Technologies 3458A User Manual page 89

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

Trigger Arm
Trigger
Event
Event
AUTO
AUTO
Any
AUTO
EXT
AUTO, EXT, TIMER,
LINE, LEVEL
AUTO
EXT
SYN
AUTO
LEVEL AUTO, EXT, TIMER,
LEVEL
AUTO
LEVEL SYN, LINE
AUTO
LINE
AUTO, EXT, TIMER,
LINE
AUTO
LINE
SYN, LEVEL
AUTO
SGL
Any
AUTO
SYN
SYN
AUTO
SYN
AUTO, EXT, LEVEL,
LINE, TIMER
EXT
AUTO
Any
EXT
EXT
AUTO, EXT, TIMER,
LINE, LEVEL
EXT
EXT
SYN
EXT
LEVEL AUTO, EXT, TIMER,
LEVEL
EXT
LEVEL SYN, LINE
1
The LEVEL event occures when the specified voltage is reached on the specified slope of the input
signal. The LEVEL trigger event or sample event can only be used for DC voltage or direct-sampled
measurements.
2
The output buffer must be empty and reading memory must be OFF or empty for the SYN event to occur.
3
The input buffer must be enabled or you must suppress cr lf when sending the TARM SGL command.
Table 21. Event Combinations
Sample
Event
One reading is taken per sample event (if the sample event is
AUTO, readings are taken continuously).
After a negative edge transition on the Ext Trig input, one
reading is taken per sample event until the specified number
of readings are completed.
Illegal
After the LEVEL event
sample event, until the specified number of readings are
completed.
Illegal
After the power line voltage crosses zero volts, one reading is
taken per sample event until the specified number of readings
are completed.
Illegal
After executing the TRIG SGL command, one reading is
taken per sample event until the specified number of readings
are completed. The trigger event then becomes HOLD. When
using the SYN sample event, the input buffer must be
enabled or you must suppress cr lf when sending the TRIG
SGL command.
After the controller requests data,
satisfied and the first reading is taken. One reading is then
taken per SYN event until the specified number of readings
are completed.
After the controller requests data,
sample event until the specified number of readings are
completed.
After a negative edge transition on the Ext Trig input, one
reading is taken per sample event until the specified number
of readings are completed.
After two negative edge transitions on the Ext Trig input, one
reading is taken per sample event until the specified number
of readings are completed.
Illegal
After a negative edge transition on the Ext Trig input followed
by the occurrence of the LEVEL event
per sample event until the specified number of readings are
completed.
Illegal
Description
1
occurs, one reading is taken per
2
both SYN events are
2
one reading is taken per
1
Chapter 4 Making Measurements
, one reading is taken
89