70 OUTPUT 722;"TARM SGL"
80 END

High-Speed DCI Example

10 OUTPUT 722;"PRESET FAST"
20 OUTPUT 722;"APER 1.4E-6"I
25
30 OUTPUT 722;"MFORMAT SINT"
40 OUTPUT 722;"MEM FIFO"
50 OUTPUT 722;"DCI 100E-3"
60 OUTPUT 722;"NRDGS 5000 AUTO"
70 OUTPUT 722;"TARM SGL"
80 END
Fast Synchronous
ACV/ACDCV Example
10 OUTPUT 722;"PRESET FAST"
20 OUTPUT 722;"MFORMAT SINT"
30 OUTPUT 722;"MEM FIFO"
40 OUTPUT 722;"SETACV SYNC"
50 OUTPUT 722;"ACV 10,2"
60 OUTPUT 722;"ACBAND 5E3,8E3"
70 OUTPUT 722;"NRDGS 20, AUTO"
80 OUTPUT 722;"TARM SGL"
90 END
Fast Random
ACV/ACDCV Example
10 OUTPUT 722;"PRESET FAST""
20 OUTPUT 722;"MFORMAT SINT
30 OUTPUT 722;"MEM FIFO"
40 OUTPUT 722;"SETACV RNDM"
50 OUTPUT 722;"ACV 10 6"
60 OUTPUT 722;"ACBAND 10E3,20E3"
70 OUTPUT 722;"NRDGS 100, AUTO"
80 OUTPUT 722;"TARM SGL"
90 END
Fast Analog ACV/ ACDCV
Example
10 OUTPUT 722;"PRESET FAST""
20 OUTPUT 722;"MFORMAT SINT
30 OUTPUT 722;"MEM FIFO"
40 OUTPUT 722;"SETACV ANA"
106
Chapter 4 Making Measurements
!TRIGGER READINGS
The following program measures DC current at the fastest possible rate.
!DCV, 10V RANGE, TARM SYN, TRIG AUTO
!LONGEST INTEGRATION TIME POSSIBLE FOR
!MAXIMUM READING RATE
!SINT MEMORY FORMAT
!ENABLE READING MEMORY
!DC CURRENT, 100mA RANGE
!5000 READINGS/TRIGGER, AUTO SAMPLE EVENT
!TRIGGER READINGS
The following program measures AC voltage using the synchronous method
at the fastest possible rate (approximately 10 readings per second). This
program can be adapted to AC+DC voltage by using the ACDCV command
instead of the ACV command in line 50.
!TARM SYN, TRIG AUTO
!SINT MEMORY FORMAT
!ENABLE READING MEMORY
!SYNCHRONOUS AC MEASUREMENT METHOD
!AC VOLTS, 10V RANGE, 2% RESOLUTION
!SIGNAL BETWEEN 5kHz AND 8kHz
!20 READINGS/TRIGGER, AUTO SAMPLE EVENT
!TRIGGER READINGS
The following program measures AC voltage using the random method at
the fastest possible rate (approximately 45 readings per second). This
program can be adapted to AC+DC voltage by using the ACDCV command
instead of the ACV command in line 50.
!TARM SYN, TRIG AUTO
!SINT MEMORY FORMAT
!ENABLE READING MEMORY
!RANDOM AC MEASUREMENT METHOD
!AC VOLTS, 10V RANGE, 6% RESOLUTION
!SIGNAL BETWEEN 10kHz AND 20kHz
!100 READINGS/TRIGGER, AUTO SAMPLE EVENT
!TRIGGER READINGS
The following program measures AC voltage using the analog method at a
fast rate. This program uses the default delay time. You can achieve faster
reading rates by specifying a shorter delay time; the resulting settling time,
however, may not produce accurate measurements. You can also achieve
unspecified faster reading rates by specifying less integration time in line 60.
This program can be adapted to AC+DC voltage by using the ACDCV
command instead of the ACV command in line 50.
!TARM SYN, TRIG AUTO
!SINT MEMORY FORMAT
!ENABLE READING MEMORY, FIFO MODE
!ANALOG AC MEASUREMENT METHOD

Hide quick links: