BEEP
BEEP
Syntax
control
Remarks
Example
CAL
CALL
Syntax
name
164
Chapter 6 Command Reference
Controls the multimeter's beeper. When enabled, the beeper emits a 1 kHz beep
if an error occurs.
BEEP [control]
The control parameter choices are:
Numeric
control
Query
Parameter
Equiv.
OFF
0
ON
1
ONCE
2
Power-on control = last programmed value.
Default control = ONCE.
•
The multimeter stores the control parameter in continuous memory (the
parameter is not lost when power is removed).
•
Query Command. The BEEP? query command returns the present beeper
mode. Refer to "Query Commands" near the front of this chapter for more
information.
•
Related Commands: TONE
OUTPUT 722;"BEEP OFF" !DISABLES THE BEEPER
This is a calibration command. Refer to the 3458 Calibration Manual for details.
Call Subprogram. Executes a previously stored subprogram.
CALL [name]
Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the SUB command for details.
Power-on name = none.
Description
Disables the beeper
Enables the beeper
Beeps once, then returns to previous mode (either OFF
or ON)