Acband - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

ACBAND

Example

ACBAND

Syntax
low_frequency
high_frequency
Remarks
158
Chapter 6 Command Reference
The time required to perform each autocal routine is:
ALL
DCV
AC
OHMS
Related Commands: CAL, SCAL, SECURE
OUTPUT 722;"ACAL ALL,3458"
AC bandwidth. Specifies the frequency content (bandwidth) of the input signal
for all AC or AC+DC measurements. Specifying the bandwidth allows the
multimeter to configure for the fastest possible measurements.
ACBAND [low_frequency][,high_frequency]
Specifies the lowest expected frequency component of the input signal.
Power-on low_frequency = 20 Hz
Default low_frequency = 20 Hz
Specifies the highest expected frequency component of the input signal.
Power-on high_frequency = 20 MHz
Default high_frequency = 2 MHz
Refer to the specifications in Appendix A for accuracy and reading rate
specifications based on the bandwidth of the input signal.
For synchronous ACV or ACDCV (SETACV SYNC command), the bandwidth
parameters are used by the multimeter to calculate time-out values and sampling
parameters. When using level triggering (default mode), if the input signal is
removed during a reading and does not return within the time limits, the
measurement method changes to random so that the reading can be completed.
(After the reading, the measurement method returns to SYNC.) For synchronous
ACV or ACDCV, it is very important that the specified bandwidth corresponds
to the frequency content of the signal being measured.
For frequency or period measurements with autorange enabled, the bandwidth
parameters are used to determine the amount of time needed for autoranging.
For these measurements, it is very important that the specified bandwidth
(especially low_frequency) corresponds to the frequency content of the signal
being measured.
If you are unsure of the frequency content of the input signal, default the
:
11 minutes
:
1 minute
:
1 minute
:
10 minutes
!RUNS ALL AUTOCALS USING
!FACTORY SECURITY CODE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents