MATH
194
Chapter 6 Command Reference
operation
The operation parameter choices are:
operation
Numeric
Parameter
Equiv.
OFF
0
CONT
1
CTHRM
3
DB
4
DBM
5
FILTER
6
FTHRM
8
NULL
9
PERC
10
PFAIL
11
RMS
12
SCALE
13
STAT
14
CTHRM2
16
K
CTHRM10
17
K
FTHRM2
18
K
FTHRM10
19
K
Description
Disables all enabled real-time math operations
Enables the previous math operation. To resume two
math operations, send MATH CONT,CONT
Result=temperature (Celsius) of a 5kW thermistor
(40653B). Function must be OHM or OHMF (10kW
range or higher).
Result = 20 x Log
(reading/REF register). The REF
10
register is initialized to 1, yielding dBV.
Result =10 x log
(reading
1O
Function must be ACV, DCV, or ACDCV.
Result = output of exponentially weighted digital
low-pass filter. Response is set by DEGREE register.
Result=temperature (Fahrenheit) of a 5kW thermistor
(40653B). Function must be OHM or OHMF (10kW
range or higher).
Result=reading-OFFSET register. The OFFSET register
is set to first reading—after that you can change it.
Result = ((reading - PERC register) / PERC register x
100.
Reading vs. MAX and MIN registers.
Result = squares reading, applies FILTER operation,
takes square root.
Result = (reading-OFFSET register) / SCALE register.
Performs statistical calculations on the present set of
readings and stores results in these registers:
SDEV = standard deviation
MEAN = average of readings
NSAMP = number of readings
UPPER = largest reading
LOWER = smallest reading
Result=temperature (Celsius) of a 2kW thermistor
(40653A). Function must be OHM or OHMF.
Result=temperature (Celsius) of a 10kW thermistor
(40653C). Function must be OHM or OHMF.
Result=temperature (Fahrenheit) of a 2kW thermistor
(40653A). Function must be OHM or OHMF.
Result=temperature (Fahrenheit) of a 10kW thermistor
(40653C). Function must be OHM or OHMF.
2
/RES register/1mW).