230
Chapter 6 Command Reference
Designates the record from which to recall readings. Records correspond to the
number of readings specified by the NRDGS command. For example, if NRDGS
specifies three readings per trigger, each record will contain three readings.
Power-on record = none.
Default record = 1.
The RMEM command automatically shuts off reading memory (MEM OFF).
This means all previously stored readings remain intact and new readings are
not stored. You can re-enable reading memory without destroying any stored
readings using the MEM CONT command.
The multimeter assigns a number to each reading in reading memory. The most
recent reading is assigned the lowest number (1) and the oldest reading has the
highest number. Numbers are always assigned in this manner regardless of
whether you're using the FIFO or LIFO mode. Records are also numbered in
this manner—the most recent record is record number 1.
When you execute the RMEM command from the front panel, readings are
copied, one at a time, to the display. After viewing the first reading, you can
view others by using the up or down arrow key. Use the left and right arrow
keys to view the reading number (left side of display) and the reading (right side
of display).
In addition to the RMEM command, you can also recall readings using the
"implied read." Refer to "Recalling Readings" in Chapter 4 for more
information.
Related Commands: MCOUNT?, MEM, MFORMAT, MSIZE, NRDGS
10 OUTPUT 722;"TARM HOLD"
20 OUTPUT 722;"DCV"
30 OUTPUT 722)"TRIG AUTO"
40 OUTPUT 722;"NRDGS 3 ,AUTO"!3 READINGS PER SAMPLE EVENT (AUTO)
50 OUTPUT 722;"MEM FIFO"
60 OUTPUT 722;"TARM SGL, 10"
70 OUTPUT 722;"RMEM 1,3,6"
80 ENTER 722;A,B,C
90 PRINT A,B,C
100 END
Request Service. Enables one or more status register conditions. When a
condition is enabled and that condition occurs, it sets the GPIB SRQ line true.
RQS [value]
value
!SUSPENDS TRIGGERING
!DC VOLTAGE MEASUREMENTS
!AUTOMATIC TRIGGERING
!ENABLES READING MEMORY, FIFO MODE
!10 GROUPS OF READINGS
!READS 1ST - 3RD READINGS of 6TH GROUP
!ENTERS READINGS INTO A, B, & C VARIABLES
!PRINTS READINGS