New Multimeter Commands

264
Chapter 7 BASIC Language for the 3458A
The following commands are not documented in chapter 6 but are included
in this supplement for your convenience. These commands will work with
all revisions of the 3458A's instrument firmware (except as noted).
ENTER user_variable Transfers a reading from the multimeter's reading
memory to a user variable. The multimeter reading is erased after execution.
Example: ENTER Dmm
OUTPUT user_variable Outputs the present value of a user variable. The
data is sent to either to the display or GPIB output buffer based on the source
from which the subprogram was executed. Example: OUTPUT Result
U_RANGE Up ranges once in the present function.
D_RANGE Down ranges once in the present function.
DSP string or user_variable Outputs to the multimeter's front-panel display
both text and user variable data (available only in REV 2.1 firmware and
greater).
DSP? Reads the present front-panel display.
SCROLL LEFT | RIGHT Scrolls the present front-panel display one
character to the left or right. This applies only to text sent with the DISP
command (for more information see chapter 6).
ECHO string Echoes the specified string back to either the multimeter's
front-panel display or GPIB. The data is sent to either to the display or GPIB
output buffer based on the source from which the subprogram was executed.
RETURN Used in a subprogram to return before the SUBEND statement.
RMATHV register , user_variable Reads a standard multimeter math
register into a user variable (available only in REV 5. 1 firmware and
greater).
WAIT msec Wait before executing the next command (32 seconds
maximum).